Open original FDA record ↗Recall record
- FDA identifier
- Z-2280-2025
- Company in FDA record
- SPINEART SA
- FDA classification date
- Aug 8, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Cannulated fenestrated polyaxial screws may be incorrectly labeled
Source facts
- Action
- On July 4, 2025, Spineart SA, issued a Urgent: Medical Device Recall Notification to affected consignees via E-Mail. On July 10, 2025, Spineart SA issued a revised notification to provide additional information. On July 16,2025, Spineart expanded the recall to include an additional product and lot. Spineart SA ask consignees to take the following steps: (1) Please acknowledge receipt of this letter by emailing to orders@spineart.us (2) Please discontinue use and return any product on hand that is listed in the Product and Distribution Information section of this letter to: Spineart U.S.A. 23332 Mill Creek Drive Suite 150 Laguna Hills, CA. 92653 (3) You may have further distributed this product; please identify concerned customers and notify them at once of this product recall by using this document. (4) Surgeons/physicians should closely monitor patients that have been implanted with these potential non-conforming screws.
- Initiated
- 2025-06-30
- Root cause
- Labeling Change Control
- Distribution
- US Nationwide distribution in the states of Florida , Kentucky, California.
- Recall number
- Z-2280-2025
- Recall status
- Open, Classified
- CDRH record ID
- 215025
- FDA classified
- 2025-08-08
- Recall event number
- 97248