Z-2280-2025: Brand Name: PERLA TL MIS Product Name: CANNULATED FENESTRATED POLYAXIAL SCREW Model/Catalog Number: MPF-

Recall record Z-2280-2025 for SPINEART SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2280-2025
Company in FDA record
SPINEART SA
FDA classification date
Aug 8, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cannulated fenestrated polyaxial screws may be incorrectly labeled

Source facts

Action
On July 4, 2025, Spineart SA, issued a Urgent: Medical Device Recall Notification to affected consignees via E-Mail. On July 10, 2025, Spineart SA issued a revised notification to provide additional information. On July 16,2025, Spineart expanded the recall to include an additional product and lot. Spineart SA ask consignees to take the following steps: (1) Please acknowledge receipt of this letter by emailing to orders@spineart.us (2) Please discontinue use and return any product on hand that is listed in the Product and Distribution Information section of this letter to: Spineart U.S.A. 23332 Mill Creek Drive Suite 150 Laguna Hills, CA. 92653 (3) You may have further distributed this product; please identify concerned customers and notify them at once of this product recall by using this document. (4) Surgeons/physicians should closely monitor patients that have been implanted with these potential non-conforming screws.
Initiated
2025-06-30
Root cause
Labeling Change Control
Distribution
US Nationwide distribution in the states of Florida , Kentucky, California.
Recall number
Z-2280-2025
Recall status
Open, Classified
CDRH record ID
215025
FDA classified
2025-08-08
Recall event number
97248

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗