Open original FDA record ↗Recall record
- FDA identifier
- Z-2281-2012
- Company in FDA record
- Advanced Neuromodulation Systems Inc.
- FDA classification date
- Aug 28, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
As of 11/30/2011, St. Jude Medical has received 110 patient complaints of warmth or heating at the implantable pulse generators (IPGs) implant site during charging for the Eon IPG and 116 reports of similar symptoms for the Eon Mini IPG, respectively.
Source facts
- Action
- ST. Jude Medical sent an Important Medical Device Correction letter dated December 19, 2011 to all affected customers. The letter identified the affected product, problem and recommendations to be followed. For questions contact your St. Jude Medical Neuromodulation Division Representative.
- Initiated
- 2011-12-19
- Root cause
- Labeling design
- Terminated
- 2015-07-24
- Distribution
- Nationwide within the US, and Internationally to Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Isreal, Italy, Japan, Mexico, Netherlands, Puerto Rico, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Recall number
- Z-2281-2012
- Recall status
- Terminated
- CDRH record ID
- 107455
- FDA classified
- 2012-08-28
- Recall event number
- 61171