Open original FDA record ↗Recall record
- FDA identifier
- Z-2283-2012
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Aug 28, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class III
Reason stated in the source
Siemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058002 and 501318002 may exhibit SLOPE or CALCV error messages (instrument errors), which indicate calibration failures.
Source facts
- Action
- SIEMENS sent an Urgent Field Safety Notice dated June 7, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter included a response form to be faxed to the recalling firm at 302-631-8467 by the consignee. If customers experienced calibration failures, they were instructed to contact the Siemens Technical Solutions Center at 800-405-6473 for assistance. Customers were also asked to forward the notification to anyone to whom they may have distributed the affected product. For questions regarding this recall call 302-631-6311.
- Initiated
- 2012-06-07
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-01-16
- Distribution
- Nationwide Distribution - USA including AK, AL, AR, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY.
- Recall number
- Z-2283-2012
- Recall status
- Terminated
- CDRH record ID
- 111478
- FDA classified
- 2012-08-28
- Recall event number
- 62712