The nurse call cable (Part Number: QDF23-0571) on the FL23SE Bed is located right under the opening of the bed head box. If an accessory (i.e. IV pole) is inserted into this slot, the circuit directing the nurse call system can be damaged.
Source facts
Action
Stryker sent an Urgent Medical Device Recall Notification letter dated August 19, 2013, to all affected consignees. The notification described the affected product, reason for recall, potential hazards, and recall customer actions and reply instructions. A Stryker Representative will contact the consignees to arrange a time to repair the unit(s) and eliminate the risk of the potential interference . Urgent questions or concerns are directed to Blaine Burnett (269)389-6954. Stryker Medical business hours are Monday-Friday 8 a.m-5 p.m. (ET).
For questions regarding this recall call 269-324-6609.
Initiated
2013-08-19
Root cause
Device Design
Terminated
2014-01-29
Distribution
Worldwide Distribution - USA including MI, WI, VA and internationally to Canada.
Recall number
Z-2283-2013
Recall status
Terminated
CDRH record ID
120936
FDA classified
2013-09-26
Recall event number
65922