Z-2283-2020: Atellica CH Enzymatic Creatinine_2, Siemens Material Number 11097533, UDI Number 00630414595795 - Product

Recall record Z-2283-2020 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2283-2020
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Jun 5, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The presence of N-acetyl-p-benzoquinone imine (NAPQI), a metabolite of acetaminophen, may cause interference in certain chemistry assays when testing patient samples.

Source facts

Action
On May 4, 2020, Siemens notified their customers that use the Fructosamine and Enzymatic Creatinine assays on the Atellica, ADVIA, and Dimension platforms of the potential for falsely depressed results in the presence of NAPQI. An Urgent Medical Device Correction Notice was sent to consignees in the US and an Urgent Field Safety Notice was sent to customers outside the US. The letter stated that the Limitations of the Procedure section of the Instructions for Use (IFU) will be updated with specific information related to the assay/platform. Customers were asked to complete and return the Effectiveness Check questionnaire attached to the letter within 30 days. In addition, customers were informed that field service personnel will be sent an Service Knowledge Base (SKB) describing the issue and instructing them how to deal with customer questions.
Initiated
2020-05-04
Root cause
Device Design
Terminated
2021-09-20
Distribution
Worldwide distribution - US Nationwide distribution including in the states of North Carolina, California, Indiana, Florida, Ohio, Arizona, Massachusetts, New Jersey, New Hampshire, Illinois, Minnesota, Arkansas, New York, Delaware, Nevada, Kentucky, Washington DC, Montana, Missouri, Colorado, Maryland, Tennessee, Oregon, Virginia, Iowa, Wisconsin, Louisiana, Georgia, Pennsylvania, North Dakota, Mississippi, South Dakota, Washington, Nebraska, Texas, Alaska, Puerto Rico, Idaho, Oregon, Maine, Vermont, West Virginia, Kansas.
Recall number
Z-2283-2020
Recall status
Terminated
CDRH record ID
181443
FDA classified
2020-06-05
Recall event number
85602

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗