Open original FDA record ↗Recall record
- FDA identifier
- Z-2284-2012
- Company in FDA record
- Steris Corporation
- FDA classification date
- Aug 28, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..
Source facts
- Action
- The firm, Steris, sent a "Voluntary Field Correction: Re-labeling of QKC1692E" letter on August 10, 2012 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to follow the steps outlined in the letter to re-label the affected QKC1692E(s) and return the completed form by September 30. 2012 to: Regulatory Affairs Specialist via Kate_Beveridge@STERIS.com or via fax at (440) 350-7082. If you have any questions, contact Director, Regulatory Affairs at 440-392-7519.
- Initiated
- 2012-08-10
- Root cause
- Device Design
- Terminated
- 2013-07-05
- Distribution
- Nationwide distribution: USA including states of: AR, AK, CA, CO, DE, FL, IL, MA, MD, NC, NE, NH, NY, OH, PA, SD, TN, TX, WI, and WV.
- Recall number
- Z-2284-2012
- Recall status
- Terminated
- CDRH record ID
- 111699
- FDA classified
- 2012-08-28
- Recall event number
- 62776