Skip to content

Public FDA record

Z-2284-2021: Monoject 0.9% Sodium Chloride Flush Syringe, 10 mL Fill, STERILE, Product Code 8881570121 The proposed

Recall record Z-2284-2021 for Cardinal Health. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2284-2021
Company in FDA record
Cardinal Health
FDA record creation date
Aug 21, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe.

Source facts

Action
On August 4, 2021 Cardinal Health issued URGENT MEDICAL DEVICE RECALL notices and response forms to customers via overnight delivery. Customers are advised to take the following actions: 1) REVIEW your inventory for affected product. 2) SEGREGATE and QUARANTINE all on-hand product per the Affected Product table. 3) COMMUNICATE the notice to clinical staff through safety huddles to ensure awareness of this product issue. 4) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to gmbfieldcorrectiveaction@cardinalhealth.com, whether or not you have affected product. 5) CONTACT the appropriate Customer Service group to arrange for return and credit of any affected product Monday Friday between 8:00am - 5:00pm EST: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other Customers888-444-5440 6) CUSTOMERS that did not receive product directly from Cardinal Health should contact the location where they purchased it. For questions related to the notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
Initiated
2021-08-04
Root cause
Device Design
Distribution
US Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.
Recall number
Z-2284-2021
Recall status
Open, Classified
CDRH record ID
188607
FDA record created
2021-08-21
Recall event number
88421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗