Skip to content

Public FDA record

Z-2286-2008: Piccolo Chemistry Analyzer software version 6.121 with Comprehensive Metabolic Lot #8111AA5 , Model Numbe

Recall record Z-2286-2008 for Abaxis Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2286-2008
Company in FDA record
Abaxis Inc
FDA classification date
Sep 17, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The product will recover tCO2 at approximately 5-7 units high relative to some controls and patient samples

Source facts

Action
The firm issued a notification letter to its consignees on April 21, 2008. The firm intends to follow up with phone calls and provide a software upgrade for the products. Contact Abaxis Customer Service at 1-510-675-6500 for assistance.
Initiated
2008-04-21
Root cause
Process control
Terminated
2011-04-26
Distribution
Product was distributed Worldwide.
Recall number
Z-2286-2008
Recall status
Terminated
CDRH record ID
71978
FDA classified
2008-09-17
Recall event number
48709

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗