Open original FDA record ↗Recall record
- FDA identifier
- Z-2287-2025
- Company in FDA record
- Spark Biomedical Inc
- FDA classification date
- Aug 11, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to manufacturing issue, neurostimulator device may experience "cable disconnect" error message and interruption to stimulation output.
Source facts
- Action
- On July 2, 2025, Spark Biomedical initiated a Urgent Medical Device Recall via phone. During the phone conversation, Spark Biomedical will ask consignees to take the following actions: 1. Spark Biomedical will replace all affected00 units at no cost. Even if the units appear to be working. 2. Current devices may continue to be used until they are replaced. 3. Please perform a check of the cable to patient controller connection before use. If this connection is loose or moves, please discontinue use of that patient controller device and quarantine that device until the replacement arrives. 4. Please notify all appropriate personnel at your facility that need to know about this recall.
- Initiated
- 2025-07-02
- Root cause
- Process control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2287-2025
- Recall status
- Open, Classified
- CDRH record ID
- 214767
- FDA classified
- 2025-08-11
- Recall event number
- 97185