During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Source facts
Action
Philips issued an IMPORTANT PRODUCT NOTICE to its consignees on 12/7/2023 via USPS mail. The notice explained the problem with the device and the circumstances under which if could occur and actions to be taken to minimize the effect of the problem. The notice further instructed the user to:
- Keep this Important Product Notice with the documentation of the system until Philips corrects
your system.
- Circulate this notice to all users of the system so that they are aware of the issue.
- Return the attached reply form to Philips to confirm that the users of the system have reviewed and understood this Important Product Notice.
For further information or support concerning this issue, please contact your local
Philips representative: Technical Support Line: 1-800-722-9377.
Initiated
2023-12-07
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-2287-2026
Recall status
Open, Classified
CDRH record ID
220332
FDA classified
2026-06-01
Recall event number
98977