Skip to content

Public FDA record

Z-2288-2018: HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 11

Recall record Z-2288-2018 for Cenorin, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2288-2018
Company in FDA record
Cenorin, LLC
FDA record creation date
Jun 26, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.

Source facts

Action
On April 9, 2018, Cenorin telephoned customers and then followed-up with a formal letter titled URGENT MEDICAL DEVICE RECALL via email. To mitigate this quality issue, Cenorin has a Corrective Kit available to replace the existing terminal blocks in the Washer/Pasteurizer as the long-term solution. In the meantime, here is how you can help with the recall process: 1. Please immediately examine your inventory for the recalled Washer/Pasteurizers listed and quarantine them to prevent further distribution. In addition, if you have further distributed this product, please notify your customers immediately using the enclosed notice. 2. If the Washer/Pasteurizers does not exhibit any of the symptoms of failure as noted, continue to use the Washer/Pasteurizers as normal until the service can be completed. If the Washer/Pasteurizers exhibit any of the symptoms of failure, immediately cease its use and contact Cenorin immediately at recall@cenorin.com or call Technical Support at 800-426-1042 ext. 224. 3. Complete the Recall Acknowledgement & Response Form and return to Cenorin as noted on form via email to recall@cenorin.com within 5 business days. Response is required. Customers with questions, please email Cenorin at recall@cenorin.com or call 253-395-2400 or toll-free 800-426-1042, extension 236 during the hours of business Monday through Friday, 7:30AM to 4:30PM, Pacific Standard Time.
Initiated
2018-04-09
Root cause
Nonconforming Material/Component
Terminated
2019-09-13
Distribution
Worldwide Distribution - US Nationwide in the states of CA, FL, IN, NC, NE, UT & WV; and countries of Malaysia and Taiwan
Recall number
Z-2288-2018
Recall status
Terminated
CDRH record ID
165073
FDA record created
2018-06-26
Recall event number
80295

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.