Skip to content

Public FDA record

Z-2288-2023: TECNIS Symfony, Extended Range of Vision IOL, REF: ZXR00, STERILE.EO, Rx Only

Recall record Z-2288-2023 for Johnson & Johnson Surgical Vision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2288-2023
Company in FDA record
Johnson & Johnson Surgical Vision, Inc.
FDA classification date
Aug 3, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential that intraocular lenses (IOLs) were manufactured with unintended toric fiducial marks.

Source facts

Action
On June 07, 20223, Johnson & Johnson Vison issued an "Urgent: Medical Device Recall" notice to affected consignees. Consignees were asked to take the following actions: 1. Identify if any of your inventory contains TECNIS IOLs from the list on pages 5-6 (see page 4 for label examples). 2. Remove all affected TECNIS IOLs from your inventory. No other TECNIS IOLs are affected by this recall. 3. Complete the attached Customer Reply Form (on page 3). We require this information for reconciliation purposes with regulatory agencies, even if you have no inventory. 4. Follow the specific instructions included in the letter "If you have product to be returned" or "If you do not have product to be returned" 5. Share this notice with anyone within your organization that needs to be informed and to any organization where the potentially affected products have been transferred. 6. If you have product complaints or adverse events to report regarding the use of the affected TECNIS IOLs, please inform Johnson & Johnson Vision. If you do report a complaint, please provide the TECNIS IOL serial number and, if a patient was involved, the date of surgery, a description of the event and patient outcome. 7. In addition, you may report any adverse events experienced or quality issues with the use of the TECNIS IOLs to the FDA's MedWatch Adverse Event Reporting program by phone at 1-800-332-1088, by fax at 1-800-332-0178, by mail at MedWatch, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787, or on the MedWatch Web site at www.fda.gov/medwatch/report.htm.
Initiated
2023-07-06
Root cause
Under Investigation by firm
Distribution
International distribution to the countries of China, India, Korea.
Recall number
Z-2288-2023
Recall status
Open, Classified
CDRH record ID
202070
FDA classified
2023-08-03
Recall event number
92666

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗