Skip to content

Public FDA record

Z-2289-2010: Zimmer ITST" INTERTROCHANTERIC/SUBTROCHANTERIC FIXATION LOCKING BOLT EXTRACTOR - NONSTERILE Zimmer, Warsa

Recall record Z-2289-2010 for Zimmer Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2289-2010
Company in FDA record
Zimmer Inc.
FDA classification date
Aug 26, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Instruments were manufactured with the incorrect grade of steel, increasing the potential for fracture during use.

Source facts

Action
Zimmer sent "Urgent: Device Removal" letters dated July 8, 2010 to consignees identifying the potential for fracture of the device. Consignees were instructed to inventory affected product and record on the Inventory Return Certification Form, to document users of the affected device on the User Facility/HCP Form, and return both forms by fax or email. Affected product is to be returned to Zimmer with a copy of the Inventory Return Certification Form. Returned product will be destroyed. Each of the affected consignees will be provided with a new instrument, manufactured with the correct 17-4 stainless steel. Consignees can contact Zimmer at 800-613-6131 or 574-372-4463
Initiated
2010-07-05
Root cause
Nonconforming Material/Component
Terminated
2011-02-25
Distribution
Worldwide Distribution in the countries of US, Canada, Spain, Germany, and Great Britain.
Recall number
Z-2289-2010
Recall status
Terminated
CDRH record ID
93239
FDA classified
2010-08-26
Recall event number
56328

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗