Open original FDA record ↗Recall record
- FDA identifier
- Z-2289-2010
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Aug 26, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Instruments were manufactured with the incorrect grade of steel, increasing the potential for fracture during use.
Source facts
- Action
- Zimmer sent "Urgent: Device Removal" letters dated July 8, 2010 to consignees identifying the potential for fracture of the device. Consignees were instructed to inventory affected product and record on the Inventory Return Certification Form, to document users of the affected device on the User Facility/HCP Form, and return both forms by fax or email. Affected product is to be returned to Zimmer with a copy of the Inventory Return Certification Form. Returned product will be destroyed. Each of the affected consignees will be provided with a new instrument, manufactured with the correct 17-4 stainless steel. Consignees can contact Zimmer at 800-613-6131 or 574-372-4463
- Initiated
- 2010-07-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-02-25
- Distribution
- Worldwide Distribution in the countries of US, Canada, Spain, Germany, and Great Britain.
- Recall number
- Z-2289-2010
- Recall status
- Terminated
- CDRH record ID
- 93239
- FDA classified
- 2010-08-26
- Recall event number
- 56328