Z-2289-2019: IMMULITE 1000 Progesterone, Model LKPW1, SMN Number 10381128

Recall record Z-2289-2019 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2289-2019
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Aug 17, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for low discordant progesterone results on a subset of patient samples. A falsely low progesterone result may lead to the consideration for additional progesterone supplementation.

Source facts

Action
Urgent Medical Device Correction notification letters were sent to customers beginning 7/10/19. Actions to be Taken by the Customer 1) Please review this letter with your Medical Director. 2) If interference is suspected, follow your established internal procedures to investigate the interference. 3) Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. 4) If you have received any complaints of illness or adverse events associated with the products, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. The firm reports that a removal at this time is not necessary because: "Siemens does not intend to remove the product as based on available data, it is estimated the occurrence is <1%. Siemens recommends that the laboratories follow their internal procedures if interference is suspected. A Follow up Urgent Medical Device Correction dated 1/13/20 was sent to customers. Actions to be Taken by the Customer Note: Interference will not be detected by quality control and the presence of the interference may not be readily identifiable. Follow your established internal procedures to determine if additional testing is needed to identify samples with suspected interference and to determine if the patient sample result is accurate. A potential approach to identify interference is to dilute the sample. In-house studies have shown that a 1:5 dilution of the sample is effective in diluting out the potential interferant. Please note that for samples with an undiluted result of d 1 ng/mL, greater variability could potentially be observed after dilution due to assay precision in this region. Instructions Please review this letter with your Medical Director. Complete and return
Initiated
2019-07-10
Root cause
Under Investigation by firm
Terminated
2024-05-28
Distribution
Global distribution. US Nationwide.
Recall number
Z-2289-2019
Recall status
Terminated
CDRH record ID
174864
FDA classified
2019-08-17
Recall event number
83349

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗