Open original FDA record ↗Recall record
- FDA identifier
- Z-2290-2016
- Company in FDA record
- Sunrise Medical (US) LLC
- FDA classification date
- Jul 25, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
There is an observed failure of the backrest bracket which can fatigue over the life time of the product. When this occurs there is the possibility of the bracket fracturing causing the backrest to detach from the chair frame.
Source facts
- Action
- The firm began calling dealers to alert them to the recall and obtain primary contacts information. Dealers were then sent their list of Quickie Q7 wheelchairs that records show being sold by them and dealers are asked to return a list of chairs they are able to locate for correction. The first dealer letter was sent via e-mail on July 7, 2016.
- Initiated
- 2016-07-01
- Root cause
- Device Design
- Terminated
- 2017-09-11
- Distribution
- US and Mexico, Japan, Chile, Kuwait, Columbia, Hong Kong, China, Uruguay, South Africa, Costa Rica, South Korea, Singapore, New Zealand, El Salvador, Bahrain, Uruguay, Lebanon, United Arab Emirates, Brazil, Argentina, Netherlands Antilles, British Virgin Islands,
- Recall number
- Z-2290-2016
- Recall status
- Terminated
- CDRH record ID
- 148019
- FDA classified
- 2016-07-25
- Recall event number
- 74596