Skip to content

Public FDA record

Z-2290-2025: Brand Name: HeartMate Product Name: Mobile Power Unit, AC Power Cord Model/Catalog Number: 107760 Softwar

Recall record Z-2290-2025 for Thoratec LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2290-2025
Company in FDA record
Thoratec LLC
FDA classification date
Aug 13, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to misaligned V-Lock feature on the Mobile Power Unit power cords, which prevents the locking mechanism from properly engaging with the Mobile Power Unit and may cause unintended disconnect.

Source facts

Action
Customer advisory letters dated June 23, 2025, will be hand delivered and by email to all impacted consignees. The customer advisory letters will be posted on the Abbott Product Advisory website for further visibility. As stated in the customer letter, affected AC Power Cords from US and OUS consignees and patient units that are experiencing connection issues will be returned and replaced. Instruction to Customers: Refer to "User Action Requested" section in attached customer letter for instructions to consignees. (Attachment D1, D2, and D3)
Initiated
2025-06-23
Root cause
Vendor change control
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, Japan.
Recall number
Z-2290-2025
Recall status
Open, Classified
CDRH record ID
214976
FDA classified
2025-08-13
Recall event number
97256

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗