Skip to content

Public FDA record

Z-2291-2008: Stryker Navigation Pin for use with the Stryker Navigation System, 4 mm x 150 mm, Stryker, Kalamazoo, MI;

Recall record Z-2291-2008 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2291-2008
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA classification date
Sep 16, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Instructions have been updated to include new warnings regarding percutaneous pin placement during surgery in order to mitigate the risk of femur fractures.

Source facts

Action
Consignees were notified via letter dated 6/5/08 providing them with additional warnings and instructions for use for these products. Contact Stryker Customer Service at 1-800-800-4236, ext. 3808.
Initiated
2008-06-05
Root cause
Labeling False and Misleading
Terminated
2009-12-01
Distribution
Nationwide, Austria, Australia, Brazil, Canada, China, Dubai, England, France, Germany, Hong Kong, India, Italy, Japan, Malaysia, Netherlands, New Zealand, Poland, Singapore, South Africa, South Korea, Sweden, Switzerland and Taiwan.
Recall number
Z-2291-2008
Recall status
Terminated
CDRH record ID
71993
FDA classified
2008-09-16
Recall event number
48710

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗