Skip to content

Public FDA record

Z-2291-2021: The Aequalis Pyrocarbon Humeral Heads. Shoulder implant component. Sold OUS only: a) Catalog # DWH039

Recall record Z-2291-2021 for Tornier S.A.S.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2291-2021
Company in FDA record
Tornier S.A.S.
FDA record creation date
Aug 13, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential patient exposure to the graphite substrate.

Source facts

Action
On June 25, 2021 Tornier sent the following notification by letter: With regards to cases where this product has already been implanted, we suggest that physicians check for the potential presence of metallic bodies in the shoulder joint and adapt patient monitoring, as appropriate, when metallic bodies are present. Otherwise, we suggest that physicians continue the monitoring of patients per their standard clinical protocols. Actions needed 1. Please circulate this Field Safety Notice to all surgeons using AEQUALIS" PYROCARBON Humeral Head Implant 2. Maintain awareness of this communication internally until all required actions have been completed within your facility. 3. Return the enclosed business reply form by email to FieldAction@wright.com within 7 days to confirm receipt of this notification. 4. Inform Tornier SAS if any of the subject devices have been distributed to other organizations. If so, provide contact details so Tornier SAS can inform the recipients appropriately. 5. Please inform Tornier SAS of any adverse event and/or report them to the Competent Authorities in accordance with current regulations and in compliance with MEDDEV 2.12 If you have any questions or concerns, please contact Customer Service +1 888 494 7950.
Initiated
2021-06-25
Root cause
Device Design
Distribution
CA, CO, CT, FL, KS, LA, MD, MI, MN, MO, NC, OR, PA, TX and WI
Recall number
Z-2291-2021
Recall status
Open, Classified
CDRH record ID
188045
FDA record created
2021-08-13
Recall event number
88174

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗