Reason stated in the source
The software calculating the patterns were found to has a rounding error for certain parameter ranges. These rounding errors could result in an incomplete cut in the donor or patient tissue.
Source facts
- Action
- WaveLight AG sent an URGENT MEDICAL DEVICE CORRECTION letter April 18, 2011 to the affected customers. The letter identified the product, problem and the action to be taken by the customer. The letter advised the customers to discontinue keratoplasty cuts with the WaveLight FS200 until the new version can be installed. Alcon will contact you to schedule a visit of a service engineer at your site, to disable the keratoplasty function in your device, until a new software version can be installed to fix this problem. If you have experienced a product problem or adverse event related to this medical device correction, please contact Alcon Medical Safety at 1-800-561-6466, option 3. Customers were requested to complete the attached reply form to acknowledge receipt of this notification and fax it to ( 817) 916-9087.
- Initiated
- 2011-04-18
- Root cause
- Other
- Terminated
- 2011-09-23
- Distribution
- Worldwide Distribution - USA including states of TX, KS, NC, CA, FL, OH, SC, MD, VA and NY and countries of Switzerland, Germany, Greece, United Kingdom, Ireland, Spain, Italy, Australia, India, Malaysia, France, Finland, Saudi Arabia, Czech Republic, Poland, China and South Korea and Iran.
- Recall number
- Z-2292-2011
- Recall status
- Terminated
- CDRH record ID
- 99665
- FDA classified
- 2011-05-25
- Recall event number
- 58601