Z-2292-2019: Luminex Aries M1 System Model: Aries M6V1 Aries Software v1.0 For In Vitro Diagnostic Use in clinical

Recall record Z-2292-2019 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2292-2019
Company in FDA record
Luminex Corporation
FDA classification date
Aug 19, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order.

Source facts

Action
On July 10, 2018, the firm sent an "URGENT: FIELD SAFETY NOTICE" to customers via email (with a follow-up telephone call) informing them that they may have received an affected barcode scanner. The Notice explained that when the potentially affected barcode scanner is used on multiple ARIES cassettes at the same time, the system may prepopulate the Sample ID field with a previously stored sample ID. This field will be overwritten by scanning or manually entering a sample, but in the event the ARIES cassette is not scanned, the prepopulated field may cause the incorrect association of sample identification to occur. The firm has conducted internal investigation that have shown that the firmware version on hand-held barcode scanner contained the incorrect version of firmware approved for use with the ARIES Systems. Given that the ARIES System barcode scanner (CN-0321-01 /JDK-2330 Rev B) has the potential to contribute to an incorrect result by creating the possibility that a Patient/Sample ID may be assigned to the wrong cassette, we are voluntarily initiating a recall to ensure no additional handheld scanners have this discrepant firmware. There have not been any adverse events associated with the use of the ARIES System barcode scanners at this time. To confirm whether or not the customer's particular barcode scanner is affected, the firm is advising its customer to verify the firmware version of the handheld barcode scanner by following the attached instructions ("Confirming the Firmware Version on the ARIES Hand-Held Barcode Scanner") . An affected unit will include the labeling JDK-2330 Rev B on the underside of the scanner AND will have firmware version PAACES00-002-R03 or CAACES00-002-R03. Please contact Technical Support if you identify a discrepant barcode scanner model. Correct Model and Firmware (below) Model/Rev. Firmware Version JDK-2330 Rev B. PAACES00-001-R03 CAACES00-001-R03 If the in
Initiated
2018-06-29
Root cause
Process control
Terminated
2021-04-28
Distribution
Worldwide distribution. US nationwide, Canada, Ireland, Netherlands, Singapore, Thailand, and United Kingdom
Recall number
Z-2292-2019
Recall status
Terminated
CDRH record ID
168673
FDA classified
2019-08-19
Recall event number
81420

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗