Z-2292-2020: ENF-GP2 Rhino-Laryngo Fiberscope - Product Usage: is indicated for use within the nasal and nasopharyngea

Recall record Z-2292-2020 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2292-2020
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Jun 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Products repaired with the incorrect service manual may be missing an O ring and are subject to rework with the corrected service manual.

Source facts

Action
On May 5, 2020, the firm distributed Urgent Medical Device Corrective Action letters to affected customers. Customers were informed that their endoscopes were incorrectly repaired and may be missing an O-ring in the battery holder. Customers were asked to take the following actions: 1. Immediately cease use of the affected endoscope. Olympus will repair your unit free of charge and provide a replacement loaner until repairs can be completed. 2. An Olympus representative will contact you to make arrangements for return of your affected endoscope. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0171 and provide your contact information as indicated in the portal. If you require additional information, please contact the firm at (484) 896-5688 from Monday till Friday 8AM ET to 5PM ET or by e-mail at laura.storms@olympus.com.
Initiated
2020-05-05
Root cause
Under Investigation by firm
Terminated
2020-09-18
Distribution
US Nationwide distribution including in the states of FL, MI, NM, NY, OH, and OR. No international distribution.
Recall number
Z-2292-2020
Recall status
Terminated
CDRH record ID
181356
FDA record created
2020-06-08
Recall event number
85621

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗