Skip to content

Public FDA record

Z-2293-2020: RadiForce GX560 monochrome LCD monitor, models 0FTD2347 and 0FTD2379 - Product Usage: is indicated for us

Recall record Z-2293-2020 for Eizo Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2293-2020
Company in FDA record
Eizo Corporation
FDA record creation date
Jun 8, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain conditions, a marble pattern infrequently appears on the monitor.

Source facts

Action
EIZO Japan announced the recall on their website on 7/30/2019. EIZO U.S. began contacting their consignees via email and phone (one as early as 8/7/2019, some on 10/18/2019, but the majority on 11/26/2019) and began updating consignee monitors with the firmware. The notification requested the consignee contact EIZO U.S. for assistance with the firmware update.
Initiated
2019-07-30
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Kansas, Louisiana, Maine, Massachusetts, Michigan, Missouri, Montana, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, South Dakota, Tennessee, Texas, Virginia, Washington, and Wisconsin. Distribution was also made to Puerto Rico. There was no government/military consignees. The countries of Canada, Australia, Austria, Belgium, Brazil, China, Czech Republic, Egypt, France, Germany Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Lebanon, Netherlands, Pakistan, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, and United Kingdom.
Recall number
Z-2293-2020
Recall status
Open, Classified
CDRH record ID
180460
FDA record created
2020-06-08
Recall event number
85299

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗