Reason stated in the source

Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

Source facts

Action
On April 20, 2026 Shimadzu Medical Systems USA issued a Urgent: Voluntary Medical Device Recall Notification via E-Mail. Shimadzu asked consignees to take the following actions: 1. If your site is on the applicable list, please schedule time with your Distributer to have your system updated. 2. Shimadzu requests that you post this letter within your facility so that all users are aware of how to avoid this issue until the corrective actions have been completed. 3. If there are any issues, please report them to your respected distributor. 4. Please complete the acknowledgement form below. 5. For additional safety measures, please see customer notification.
Initiated
2026-04-20
Root cause
Component design/selection
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, The Bahamas, China, Denmark, Haiti.
Recall number
Z-2294-2026
Recall status
Open, Classified
CDRH record ID
220078
FDA classified
2026-06-02
Recall event number
98571

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗