Open original FDA record ↗Recall record
- FDA identifier
- Z-2295-2010
- Company in FDA record
- Biomerieux Inc
- FDA classification date
- Aug 27, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Foil pouch was punctured by drug susceptibility cards during packaging, allowing moisture into pouch, resulting in inaccurate and/or incorrect drug susceptibility results.
Source facts
- Action
- BioMerieux issued an Urgent Product Removal Notice, dated July 29, 2010 to consignees, identifying the reason for the recall and the affected lots. Customers were instructed to discontinue use of and destroy any remaining inventory of the affected lot, and to notify their customers who received the product. All consignees were instructed to complete and return the Customer Acknowledgement and Product Replacement Form. These cards will be used to distribute replacement cards. Consignees can contact BioMerieux at 1-314- 731-8667.
- Initiated
- 2010-07-29
- Root cause
- Packaging
- Terminated
- 2010-11-02
- Distribution
- Worldwide Distribution: Country of Canada only. No product was distributed in the US.
- Recall number
- Z-2295-2010
- Recall status
- Terminated
- CDRH record ID
- 93305
- FDA classified
- 2010-08-27
- Recall event number
- 56365