Z-2295-2019: Brand Name Burn Navigator Model Numbers: Catalog 1120, Burn Navigator H2 Catalog 1125, Burn Navigator H2

Recall record Z-2295-2019 for Arcos, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2295-2019
Company in FDA record
Arcos, Inc
FDA record creation date
Aug 21, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Battery packs may heat up abnormally and ignite in rare instance.

Source facts

Action
On March 20, 2015, the firm sent an email with a Battery Replacement Program Attachment. The Attachment informed customers that it main hardware supplier had notified them that some batteries used with tablet computers have in rare instances, overheated and caught fire. The Attachment instructed customers to check the batteries in the computer tablets to determine if they matched the serial numbers for the batteries being replaced. Customer were instructed to: 1) Remove Battery Pack - shut down their PC and remove the battery pack (noting that there are two battery slots, one on each the left and right side of the Tablet Computer. 2) Review Battery Pack Model Number and Manufacturing Lot Number - by reviewing battery pack nameplate. The following Battery Pack Model Numbers and Manufacturing Lot Numbers must be replaced: Battery Pack Model Numbers: Manufacturing Lot Number CF-VZSU53AW Beginning with B, C1, C2, C3, C4, C5, CF-VZSU53AR C6, C7, C8, C93, C94, C96, C9C, C9J, CF-VZSU53AQ C9K If the customer has the affected batteries, they are to contact the firm whether or not the batteries match the criteria so that the firm will know the status of the customer's batteries. If batteries match the criteria, the firm will replace them for free. Contact Information: Toll Free Phone Number: 877-542-8025 In addition Tablet Computers can continue to be operated with wall power or with batteries not affected by the recall.
Initiated
2015-03-16
Root cause
Component design/selection
Terminated
2022-02-09
Distribution
US: AZ, LA, OR, MD, TX, and WA. OUS: No products were distributed outside the US.
Recall number
Z-2295-2019
Recall status
Terminated
CDRH record ID
169028
FDA record created
2019-08-21
Recall event number
81594

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗