If the bed is in the upright position and the bar is dropped rapidly or in free-fall, it could strike the patient in the bed, and the drop could present a laceration hazard by creating a sharp edge on the metal bar.
Source facts
Action
Consignees were notified via recall letter dated 7/7/08 to discontinue use of these devices until the firm visits and replaces the devices. Contact Zimmer at 1-800-321-5533 for assistance.
Initiated
2008-07-07
Root cause
Device Design
Terminated
2009-11-12
Distribution
Nationwide and Canada.
Recall number
Z-2296-2008
Recall status
Terminated
CDRH record ID
71983
FDA classified
2008-09-16
Recall event number
48714