Z-2296-2024: TOTAL KNEE - 297835-Procedure Kit Catalog Number: RGTK10K

Recall record Z-2296-2024 for American Contract Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2296-2024
Company in FDA record
American Contract Systems Inc
FDA classification date
Jul 5, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

Source facts

Action
American Contract Systems issued Urgent Voluntary Recall Letter (FA-2024-018) via email on 5/15/24 to Distributors. Distributors to notify end-users. Letter states reason for recall, health risk and action to take: Immediately discontinue use of the cast padding component (DISCARD ITEM 4927009046-CAST PADDING, 6" Cotton Blend Rayon) included in each kit/tray identified in ATTACHMENT 1. Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing labeling. "Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. "Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Complete the enclosed Response Form and return as soon as possible. "If credit is requested, please send an email to quality@owens-minor.com.
Initiated
2024-05-15
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2296-2024
Recall status
Open, Classified
CDRH record ID
208237
FDA classified
2024-07-05
Recall event number
94765

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗