Skip to content

Public FDA record

Z-2296-2026: Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta In

Recall record Z-2296-2026 for Bayer Medical Care, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2296-2026
Company in FDA record
Bayer Medical Care, Inc.
FDA classification date
Jun 3, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.

Source facts

Action
On April 24, 2026, URGENT MEDICAL DEVICE RECALL letters were sent to customers. Customers were instructed to immediately halt the use of the affected product and review inventory for impacted product and quarantine. A Return Goods Authorization will be sent to return any remaining affected inventory. Place the affected product in a box and label the outside of the container with the RGA number provided and RECALL ID: SA-2026-RAD-03 in large, bold writing. Use the shipping label provided via email when you receive your RGA number. Please return the affected product as soon as possible. Please distribute this letter to other appropriate departments and personnel within your facility who may need awareness of this recall.
Initiated
2026-04-24
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.
Recall number
Z-2296-2026
Recall status
Open, Classified
CDRH record ID
220075
FDA classified
2026-06-03
Recall event number
98881

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗