Open original FDA record ↗Recall record
- FDA identifier
- Z-2297-2024
- Company in FDA record
- American Contract Systems Inc
- FDA classification date
- Jul 5, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.
Source facts
- Action
- American Contract Systems issued Urgent Voluntary Recall Letter (FA-2024-018) via email on 5/15/24 to Distributors. Distributors to notify end-users. Letter states reason for recall, health risk and action to take: Immediately discontinue use of the cast padding component (DISCARD ITEM 4927009046-CAST PADDING, 6" Cotton Blend Rayon) included in each kit/tray identified in ATTACHMENT 1. Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing labeling. "Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. "Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. "Complete the enclosed Response Form and return as soon as possible. "If credit is requested, please send an email to quality@owens-minor.com.
- Initiated
- 2024-05-15
- Root cause
- Under Investigation by firm
- Distribution
- US Nationwide distribution.
- Recall number
- Z-2297-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208238
- FDA classified
- 2024-07-05
- Recall event number
- 94765