Z-2298-2016: Syva EMIT 2000 Theophylline - Catalog # 4P019UL /SMN#10445324 is a homogeneous enzyme immunoassay intende

Recall record Z-2298-2016 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2298-2016
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Jul 28, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens Healthcare Diagnostics has confirmed the Syva EMIT 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for commercially available Bio-Rad Liquichek TDM Quality Control Levels and patient samples.

Source facts

Action
Siemens Healthcare Diagnostics mailed a letter to customers on June 17, 2016, asking that they discontinue use of and discard the affected kit lots identified in the recall notification letter. Customers were also asked to complete and return the Field Correction Effectiveness Check/Product Replacement Form that was mailed with the Recall Notification Letter.
Initiated
2016-06-17
Root cause
Process control
Terminated
2017-03-15
Distribution
Distributed to: CA, CT, DE, FL, GA, IA, IN, KS, KY, LA, MA, MD, MI, MO, NC, NJ, NM, NY, OH, OK, PA, TN, VA, WA
Recall number
Z-2298-2016
Recall status
Terminated
CDRH record ID
147833
FDA classified
2016-07-28
Recall event number
74524

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗