Skip to content

Public FDA record

Z-2299-2020: Triever20 Aspiration Guide Catheter a component of FlowTriever Retrieval/ Aspiration System, UDI: 0085029

Recall record Z-2299-2020 for INARI MEDICAL. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2299-2020
Company in FDA record
INARI MEDICAL
FDA record creation date
Jun 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The hemostasis valve of aspiration guide catheter devices may fail to seal resulting in a loss of vacuum during thrombus aspiration, which may result the procedure being prolonged as another device must be used or suboptimal thrombus removal if the problem is not detected.

Source facts

Action
On 03/23/20, Urgent Medical Device Recall notices were emailed to customers. Customers were asked to take the following actions: 1.Immediately review your inventory for the affected devices and quarantined to prevent their use. 2.Complete and return the Customer Response Form. 3.Return quarantined devices to recalling firm. 4.Report any adverse health consequences experienced with the use of this product to the recalling firm. Customers with additional questions were encouraged to contact the firm at: 877.923.4747 or customerservice@inarimedical.com Additional recall notices with an expanded scope, dated 03/26/20, were also sent to additional customers.
Initiated
2020-03-23
Root cause
Equipment maintenance
Terminated
2021-08-13
Distribution
U.S Distribution in the states of .: IA, GA, MO, MI, AZ, FL, NJ, MN, LA, PA, AL, OH, NY, WI, CA, TX, TN, NH, IL, NC, CT, CO, OK, OR, IN, MA, VA, KY, WA, MS, SC, UT, SD, ND, AR, ID, DE, including DC.
Recall number
Z-2299-2020
Recall status
Terminated
CDRH record ID
180814
FDA record created
2020-06-09
Recall event number
85424

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗