Z-2301-2018: OmniDiagnost Classic X-ray system Angiographic and Diagnostic, Product Nos. 70859, 708023, 708024, 70802

Recall record Z-2301-2018 for Philips Medical Systems Nederlands. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2301-2018
Company in FDA record
Philips Medical Systems Nederlands
FDA classification date
Jun 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user. This could potentially lead to the patient falling off the table. There is a risk of injury for the patient, user, or bystander.

Source facts

Action
On May 17, 2018, the firm issued Urgent Medical Device Correction letters to affected consignees. Customers were advised of the product issue. Customers were instructed to do the following: Stop using the system and call your Philips representative if any of the following situations occurs: . the table height or tilt movement is not working. . you notice a cracking sound or snapping sound different than the normal sound during height or tilt movement. . you experience a blockade in the table height or tilt movement during system movements. . the system is having an unexpected collision or the system had an unexpected collision in the past 2 months. . you have used a CPR stand that obstructed the system, or . you notice any (other) unusual system behavior beyond the normal use of the system. Avoid collisions with the table and base stand of the system by ensuring that no obstacles are placed around the system. Customer shall ensure that all staff with access to the affected systems are informed of the content of this Field Safety Notice. A copy of this Field Safety Notice shall be placed together with the documentation of the system until the system has been corrected by Philips. The firm plans to correct all potentially affected products in the field by means of a field change order free of charge. Customers will be contacted by a local Philips representative to schedule the corrective action, which will start effective July 2018. Customer with further questions may contact Philips Technical Support Line: 1-800-722-9377.
Initiated
2018-05-01
Root cause
Device Design
Terminated
2020-06-08
Distribution
US nationwide distribution.
Recall number
Z-2301-2018
Recall status
Terminated
CDRH record ID
165012
FDA classified
2018-06-27
Recall event number
80271

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗