Open original FDA record ↗Recall record
- FDA identifier
- Z-2301-2026
- Company in FDA record
- CareFusion 303, Inc.
- FDA classification date
- Jun 3, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fingerprint scanner may overheat to a temperature to cause 1st degree burn.
Source facts
- Action
- On April 30, 2026, CareFusion (subsidiary of Becton, Dickinson and Company) issued a Urgent Medical Device Recall Notification, via mail or Email. CareFusion ask consignees to take the following actions: 1. Dispensing cabinets can continue to be used as intended. 2. If you suspect that the Bio-ID Scanner is hot, please discontinue the use of Scanner and file a complaint. 3. Review and update facility policies and procedures related to situations where the Bio-ID Scanner is unavailable to mitigate delays in care. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form.
- Initiated
- 2026-04-30
- Root cause
- Component design/selection
- Distribution
- US: Nationwide OUS: AE, AU, BE, BH, BR, CA, CL, ES, GB, IT, KW, MA, MX, NZ, PT, QA, SA, SG, TH, TW
- Recall number
- Z-2301-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220101
- FDA classified
- 2026-06-03
- Recall event number
- 98895