Skip to content

Public FDA record

Z-2302-2018: Flexible Smoothbore Breathing Circuit with Exhalation Port 22mm, Product No. 80062000

Recall record Z-2302-2018 for Intersurgical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2302-2018
Company in FDA record
Intersurgical Inc
FDA record creation date
Jun 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When assembling an HME or Filter on this breathing circuit the tubing may be pushed up and inadvertently cover or block the exhalation port. This could potentially pose a serious health risk to the patient.

Source facts

Action
On April 25, 2018, the firm issued an Urgent Medical Device Recall letter to its distributors. Customers were advised of the product issue and recall. Recall Instructions: 1. Immediately discontinue use and quarantine any stock with the product number B0062000 described above in this recall notification. 2. If you have distributed any of this product, please immediately contact your customers, advise them of the recall situation and have them return their outstanding recalled stock to you. 3. Contact Intersurgical Customer Support at 800-828-9633 for a Return Goods Authorization Number (RGA). Hours of operation are Monday through Friday, 8:00 a.m. to 4:30 p.m. EST. 4. Once you have received the RGA Number, please enter it in the space provided on the Customer Product Recall Acknowledgement Form: a. Return the completed form via email to Intersurgical Inc. at Support@intersurgicalinc.com. b. The subject line should contain your Company name and RGA number. i. This will allow us to document your receipt of this letter and the amount of product you have on-hand for return. 5. Intersurgical will credit your account when the product is received. If you have any other questions, feel free to contact Intersurgical Customer Support at 1-800-828-9633.
Initiated
2018-04-25
Root cause
Device Design
Terminated
2019-04-09
Distribution
Distributors in IL, NJ, NV, TX, UT.
Recall number
Z-2302-2018
Recall status
Terminated
CDRH record ID
164961
FDA record created
2018-06-27
Recall event number
80250

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗