Z-2302-2020: Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Ar

Recall record Z-2302-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2302-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Jun 9, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated and performed on an alternative x-ray system

Source facts

Action
Siemens Healthineers issued Urgent Medical Device Correction via AX036/19/S, dated 4/30/20, letter states reason for recall, health risk, and action to take: Siemens will exchange the affected power distributor of the Multi Display Manager. We advise you not to move the Multi Display Manager Container or to open the power distributor before the corrective action can be completed. A service organization will contact you to schedule an appointment to perform the correction via Update Instruction AX035/19/S. Please feel free to contact service organization at 1-800-888-7436 for an earlier appointment. Following the correction, the cause will be eliminated, and any recurrence of this potential fault is prevented; Acknowledge Receipt of this Safety Advisory Notice, By signing this document, you are acknowledging that you have read and understand the content therein.
Initiated
2020-04-30
Root cause
Nonconforming Material/Component
Terminated
2020-09-11
Distribution
US Nationwide distribution.
Recall number
Z-2302-2020
Recall status
Terminated
CDRH record ID
181459
FDA record created
2020-06-09
Recall event number
85647

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗