Open original FDA record ↗Recall record
- FDA identifier
- Z-2303-2013
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Sep 30, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.
Source facts
- Action
- On 11/07/2012, the firm sent "Urgent Field Safety Notices to their customers. The product issue was described and recommended actions were provided. Customers may contact 1 (800) 736-0600, option 4 if questions.
- Initiated
- 2012-10-08
- Root cause
- Process control
- Terminated
- 2013-12-24
- Distribution
- Nationwide Distribution, including the states of: AK, AL, AR, AZ, BM, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV.
- Recall number
- Z-2303-2013
- Recall status
- Terminated
- CDRH record ID
- 121492
- FDA classified
- 2013-09-30
- Recall event number
- 66108