- FDA identifier
- Z-2305-2012
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Aug 31, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
When using the syngo Quality Control package the system may not perform as intended for Multi-Rule QC violation [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions on page 2 of the recall notification. QC status may not be flagged appropriately. If auto-verification rules are defined to hold patient results when QC status is out, the patient results will not be held as expecte
Source facts
- Action
- Siemens sent a Urgent Field Safety Notice letter dated May, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed the customers to re-set the Multi-Rule option and instructions are provided on how to do so. This recall notification also included a Field Safety Effectiveness check form to please fax to the Siemens Technical Solutions Center at 302-631-8467 BY June 15th.
- Initiated
- 2012-05-29
- Root cause
- Software design
- Terminated
- 2014-04-10
- Distribution
- USA ( nationwide ) including the states of AZ, CA, FL, IL, KY, MD, MA, MN, MS, MO, NE, NH, NJ, NY, NC, OH, PA, TX, VA and WV.
- Recall number
- Z-2305-2012
- Recall status
- Terminated
- CDRH record ID
- 111471
- FDA classified
- 2012-08-31
- Recall event number
- 62709