Skip to content

Public FDA record

Z-2306-2021: EPIQ Ultrasound Systems, Model: EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CV

Recall record Z-2306-2021 for Philips Ultrasound Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2306-2021
Company in FDA record
Philips Ultrasound Inc
FDA record creation date
Aug 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ultrasound system control panel arm swivel lock could fail, which could cause the control panel and monitor to swivel freely while a user is positioning the monitor, or transporting/moving the system, which could lead to patient or user musculoskeletal injury.

Source facts

Action
Medical device correction notices were mailed to customers who were asked to take the following actions to prevent risks for patients or users: When Positioning or Transporting/Moving the EPIQ Ultrasound System: A) Prior to positioning or transporting/moving the system, ensure the control panel locking mechanism is engaged to prevent the control panel and monitor from swiveling freely. B) To ensure the control panel locking mechanism is working properly: 1. Engage the system brake. 2. Observe the lock/unlock button on the system handle. It should not be illuminated. 3. If it is illuminated press button to lock arm. Illumination should cease and arm should lock. 4. Using the system handle, attempt to rotate the control panel. If the control panel does not rotate, the system is locked and ready to be transported/moved. 5. Release the brake and transport/move the system as needed. C) Do not make any twisting or sudden motions with the control panel when adjusting or steering the system. D) When transporting/moving the system, push from behind using the back handles on the control panel. Additional information can be obtained from your local service representative: 1-800-722-9377 Complete and return the response form.
Initiated
2021-07-22
Root cause
Device Design
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, UT, CA, CO, WA, MS, NC, NV, GA, LA, VA, SC, NE, TN, HI, OR, ID, MI, OH, IN, IL, OK, TX, PA, NJ, CT, NY, DE, SD, WI, VT, MN, KS, MO, KY, MA, NH, IA, MT, NM, ME, AZ, AR, AL, WY, ND, AK, MD, RI, WV, DC, VI and the countries of Argentina, Australia, Bahrain, Peru, Serbia, Bolivia, Moldova, Albania, Ukraine, Uruguay, Belarus, Hungary, Greece, Honduras, Montenegro, Senegal, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Cyprus, Dominican Republic, Germany, Ecuador, Estonia, Finland, Taiwan, Georgia, United Kingdom, Guatemala, Hong Kong, Indonesia, Iran, Israel, Yemen, Jordan, Korea, Republic of, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malta, Morocco, Mauritius, Nicaragua, New Zealand, Norway, Oman, United Arab Emirates, Austria, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Thailand, Czech Republic, Tunisia, Uzbekistan, South Africa, Sweden, Switzerland, Netherlands, France, Monaco, Martinique, New Caledonia, Italy, China, India, Denmark, Singapore, Belgium, Spain, Mexico, Viet Nam, Philippines, Malaysia, Japan, Bangladesh, Ireland, Luxembourg, Andorra, Nepal, Myanmar, Pakistan, Brunei Darussalam, French Guiana, Sri Lanka, Kazakhstan, Egypt, Turkey, Kuwait, Romania, Slovakia, Croatia, Slovenia, Bosnia and Herzegovina.
Recall number
Z-2306-2021
Recall status
Completed
CDRH record ID
188369
FDA record created
2021-08-16
Recall event number
88313

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗