Skip to content

Public FDA record

Z-2307-2008: NeuViz Dual series computed Tomography Scanner System, X-Ray System

Recall record Z-2307-2008 for Philips And Neusoft Medical Systems Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2307-2008
Company in FDA record
Philips And Neusoft Medical Systems Co., Ltd.
FDA classification date
Sep 17, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential for R-host box and components inside to be detached of its mounting during the gantry rotor rotation. There is a potential for components to be damaged inside the system gantry and for the internal components to be expelled from the system gantry and injure the patient, operator, bystander or service people.

Source facts

Action
On 6/23/08/08, the Philips Neusoft sent a letter dated 6/10/08 to their US customers explaining the problem, under what circumstances it can occur, and the actions that should be taken by the customer/user in order to prevent risks for patients or users. The field correction is install a R-host bracket to aid supporting of the R-host computer. The letter included additional information to customers who had already received the field action (this was an on-going activity that was re-classified as a potential safety). Customers did not receive the additional information about the status of their upgrade if the upgrade had not occurred at the time of the mailing. They were informed that they would be contacted by Newsoft Medical systems representative for the planning of this repair. If you have questions contact Service Support Department at nms-service@neusoft.com.
Initiated
2008-06-23
Root cause
Nonconforming Material/Component
Terminated
2010-09-24
Distribution
The units were distributed to one distributor in NC, and further distributed to medical facilities located in CO, FL, KY, NC, SC, TN and TX.
Recall number
Z-2307-2008
Recall status
Terminated
CDRH record ID
72008
FDA classified
2008-09-17
Recall event number
48727

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗