Skip to content

Public FDA record

Z-2308-2019: BD Vacutainer LH (Lithium Heparin) 34 I.U. Plus Blood Collection Tubes, Catalog Number 368494

Recall record Z-2308-2019 for Becton Dickinson & Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2308-2019
Company in FDA record
Becton Dickinson & Company
FDA record creation date
Aug 21, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

False elevation of carboxyhemoglobin (COHb) results have been reported when using venous blood gas samples collected with tubes analyzed with the IL GEM 4000 instrument.

Source facts

Action
Urgent Medical Device Correction notification letters dated 6/12/19 were sent to customers.
Initiated
2019-06-12
Root cause
Device Design
Terminated
2020-06-29
Distribution
Worldwide Distribution, including US Nationwide
Recall number
Z-2308-2019
Recall status
Terminated
CDRH record ID
173993
FDA record created
2019-08-21
Recall event number
82963

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗