Z-2308-2021: Model L43K Intraoperative Probe-For the diagnostic ultrasound evaluation during robotic/non-robotic intra

Recall record Z-2308-2021 for Hitachi Healthcare Americas Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2308-2021
Company in FDA record
Hitachi Healthcare Americas Corporation
FDA record creation date
Aug 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The screw cover at the tip of the probe may fall off after a cleaning and sterilization process. Users advised to conduct a preventive inspection of the probe before and after sterilization/disinfection/cleaning and before and after an interoperative procedure.

Source facts

Action
Hitachi issued Urgent Medical Device Correction Letter to customers on 6/30/2021 using FedEx (with tracking). Letter states reason for recall, health risk and action to take: Customers/Users are advised to: 1. Remember to thoroughly inspect the probe before and upon completion of your cleaning and sterilization process and before and after any interoperative procedure as indicated in the Cleaning, Disinfection and Sterilization Instruction Manual including verifying the 2 screw covers are not loose, damaged, dislodged, or missing. 2. If any abnormalities exist, immediately remove the probe from service and contact Hitachi Customer Service at 1-800-800-4925. 3. Please complete the attached Medical Device Recall Return Response Form and return to Hitachi at your earliest convenience. Type of Action by the Company: An updated manual is included with newly manufactured L43K probes to include the above diagram for additional clarification. (7) Other Information: For questions, comments, or concerns, please contact: Aaron Pierce Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749 Email piercea@hitachihealthcare.com
Initiated
2021-06-30
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-2308-2021
Recall status
Open, Classified
CDRH record ID
188516
FDA record created
2021-08-16
Recall event number
88367

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗