Due to incorrect functional length on device labeling.
Source facts
Action
On 04/03/2026, the firm emailed a "RECALL NOTICE" to it single customer informing them that the firm determined the pistons were manufactured with an incorrect functional length. Devices are labeled as 3.75 mm but have an actual functional length of 4.00 mm.
The customer was instructed to:
" Immediately discontinue use and quarantine affected product
" Return affected product using the provided RMA
" If any devices have been used, provide the date(s) of surgery on the response form
" Complete and return the enclosed Recall Response Form
For questions - contact Grace Medical at 901-386-0990 or Thomas.fearnley@gracemedical.com.
Initiated
2026-04-03
Root cause
Process control
Distribution
U.S Nationwide distribution in the state of MD.
Recall number
Z-2308-2026
Recall status
Open, Classified
CDRH record ID
219759
FDA classified
2026-06-04
Recall event number
98773