Open original FDA record ↗Recall record
- FDA identifier
- Z-2309-2012
- Company in FDA record
- Cook Ireland Ltd.
- FDA classification date
- Sep 1, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Cook Ireland received a product complaint from Japan on December 22nd, 2011. The complainant reported that a 5 French Zilver Vascular Stent (part number ZIV5-18-125-8-80 from Lot# CF694429) was inserted into a sheath over a wire guide where it became lodged in the sheath hub. After further examination, the user determined that the product was a 6 French Zilver Vascular Stent (part number ZIV6-35
Source facts
- Action
- The firm, Cook Medical, sent a letter dated February 13, 2012 to foreign consignee. The letter described the product, problem and actions taken. The firm stated that re-training of the relevant personnel has been carried out; they will continue to monitor to identify potential emerging trends and further corrective and preventive action will be considered in an effort to avoid the recurrence of similar events. If you have any question contact in US: Customer Relations Manager at 800.346.2686.ext 2222 or email: rita.harden@cookmedical.com; OUS: Customer Quality Supervisor at +353 61 239343.
- Initiated
- 2012-01-05
- Root cause
- Packaging process control
- Terminated
- 2013-02-04
- Distribution
- Worldwide distribution: USA (nationwide) and country of: Japan.
- Recall number
- Z-2309-2012
- Recall status
- Terminated
- CDRH record ID
- 108136
- FDA classified
- 2012-09-01
- Recall event number
- 61451