The firm identified a precipitation issue which may cause late Ct reactors (a false positive result) with a rate of 5 to 10% for affected vials of RNA Master Mix.
Source facts
Action
/The firm disseminated a technical bulletin beginning on 06/25/2021 by email identifying the /problem and providing additional instructions to be implemented when using the product.
Initiated
2021-06-25
Root cause
Nonconforming Material/Component
Distribution
Maine, New York, Alabama, Pennsylvania, Florida, Illinois, New Jersey, Connecticut, Georgia, Puerto Rico, Costa Rica, Hong Kong, Philippines, Barbados
Recall number
Z-2310-2021
Recall status
Open, Classified
CDRH record ID
188443
FDA record created
2021-08-17
Recall event number
88344
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.