Open original FDA record ↗Recall record
- FDA identifier
- Z-2311-2026
- Company in FDA record
- ICU Medical, Inc.
- FDA classification date
- Jun 5, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Source facts
- Action
- On April 29, 2026, the firm notified consignees via letters titled "URGEND MEDICAL DEVICE CORRECTION." Customers were instructed to identify all affected pumps in their possession and ensure all users or potential users of the pumps are immediately made aware of the notification. During bolus programming, clinicians should check whether the soft limit override icon is displayed next to the bolus dose entry field on the programming screen. If the pump displays the icon, the clinician should review the program and make any necessary adjustments before confirming the program. The firm is developing a software update to correct the issue and will notify customers when it is available.
- Initiated
- 2026-04-29
- Root cause
- Software Design Change
- Distribution
- Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
- Recall number
- Z-2311-2026
- Recall status
- Open, Classified
- CDRH record ID
- 220055
- FDA classified
- 2026-06-05
- Recall event number
- 98872