St. Jude Medical has initiated a recall of one lot of Livewire TC" Ablation Catheter. The product was mis-labelled as having an 8 mm tip instead of 4 mm tip. This could potentially result in elevated temperatures and formation of coagulum at the tip electrode. If this coagulum were to detach from the product, the worst case this could potentially result in embolization.
Source facts
Action
Consignees were sent a "St Jude Medical Urgent Product Recall" letter addressed to "Dear Doctor" and a "Doctor's Acknowledgement Form" . The letter described the product, problem and the possibility of a health hazard. They also requested discontinue of product and gave instructions for return of product.
Contact St. Jude Medical at 952-933-4700 if you have questions.
Initiated
2008-06-05
Root cause
Labeling mix-ups
Terminated
2012-02-04
Distribution
Canada and Belgium
Recall number
Z-2313-2008
Recall status
Terminated
CDRH record ID
72033
FDA classified
2008-09-20
Recall event number
48740