Open original FDA record ↗Recall record
- FDA identifier
- Z-2315-2010
- Company in FDA record
- AdvanDx, Inc.
- FDA classification date
- Aug 31, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some strains of streptococus anginosus cross-reach with E. faecalis/OE PNA FISH to produce a false positive green signal.
Source facts
- Action
- The firm, AdvanDx, sent a letter dated June 8, 2010, and email notification on June 11, 2010, to all customers. The letter described the product, problem and action to be taken by customers. AdvanDx notified the customers that there was a change to the package insert under "limitations" to include the statement "some strains of Streptococcus anginosus produce a false green positive fluorescence due to sequence similarities and enclosed a package insert for all E. facealis/OE PNA FISH (KT003) kits. The customers were instructed to replace the Package Insert (revision B) included with any kits in their inventory with the enclosed Package Insert (revision C). Note: All shipments of E. faecalis/OE PNA FISH, as of June 14, 2010, will contain this new package insert. If you have any questions or requests for further information, please contact AdvanDx Technical Support at US (Toll Free): 1-800-376-0009 or email techsupport@advandx.com.
- Initiated
- 2010-06-08
- Root cause
- Device Design
- Terminated
- 2011-06-06
- Distribution
- Worldwide distribution: USA and countries including Danmark, Finland and Sverige (European).
- Recall number
- Z-2315-2010
- Recall status
- Terminated
- CDRH record ID
- 92578
- FDA classified
- 2010-08-31
- Recall event number
- 56064