Skip to content

Public FDA record

Z-2316-2018: MX-10 surveillance laser system Product Usage: The expected usage of the product is to facilitate observ

Recall record Z-2316-2018 for Wescam Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2316-2018
Company in FDA record
Wescam Inc
FDA record creation date
Jul 15, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Product has a nominal ocular hazard distance (NOHD) that is longer than specified and persons could be exposed to light in excess of the exposure limit.

Source facts

Action
Firm sent a Service Bulletin dated June 2018 to customers. The Bulletin identified the affected product, problem and actions to be taken. For question call 888-5-WESCAM or email support.wescam@L3T.com.
Initiated
2018-05-08
Root cause
Radiation Control for Health and Safety Act
Terminated
2021-08-24
Distribution
US Nationwide Distribution
Recall number
Z-2316-2018
Recall status
Terminated
CDRH record ID
165340
FDA record created
2018-07-15
Recall event number
80392

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗