Open original FDA record ↗Recall record
- FDA identifier
- Z-2316-2021
- Company in FDA record
- Philips North America Llc
- FDA record creation date
- Aug 18, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When setting patient weight unit preferences to pounds, SynchRight P3T (Personalized Patient Protocol) software will result in increased contrast volume, resulting in increased contrast volume recommended for the patient.
Source facts
- Action
- Philips issued Urgent Medical Device Correction letter dated 6/8/21 to consignees via certified mail. Letter states reason for recall, health risk and action to take: following steps: Go to Preferences> Regional Settings > Weight units, select the Weight units to Kg. Step 1: Verify that the patient weight unit is set to Kg, rather than Lbs. Step 2: Change the weight unit to Kg if weight unit is set to Lbs, referring to the screenshot below Step 3: Log out and then log back in to make the changes effective. Please share this notice with all those who need to be aware within your organization. Please complete and return the attached response form to Philips Healthcare promptly upon receipt and no later than 30 days from receipt, via email to CTNM.QARA@PHILIPS.COM The software update is planned to be released in the 3rd quarter of 2021. Once the software update is released, a Philips representative will contact you to schedule a software update to permanently resolve the issue. For North America, contact the Customer Care Solutions Center (1-800-722-9377).
- Initiated
- 2021-06-08
- Root cause
- Software design
- Terminated
- 2023-08-10
- Distribution
- US Nationwide distribution in the states of AZ, FL, LA, MN, NE, NV, NY, PA.
- Recall number
- Z-2316-2021
- Recall status
- Terminated
- CDRH record ID
- 188422
- FDA record created
- 2021-08-18
- Recall event number
- 88336