Z-2317-2017: DePuy Synthes various orthopedic instruments modified by U.S. Distributors These instruments are used in

Recall record Z-2317-2017 for DePuy Orthopaedics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2317-2017
Company in FDA record
DePuy Orthopaedics, Inc.
FDA record creation date
Jun 8, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Products were made outside of Quality System Regulation, and potentially outside of premarket submission (510k/PMA) for certain devices. The safety or effectiveness of these devices cannot be assured.

Source facts

Action
DePuy Synthes sent an URGENT INFORMATION RECALL NOTICE FOR INSTRUMENTS MODIFIED BY U.S. DISTRIBUTORS dated January 6, 2017, to all affected customers. . Customerss were instructed to identify all medical facilities that may have used or received the affected instruments and identify the modified instruments used at each facility. This information was to be used to generate Reconciliation Forms for each impacted Medical Facility. The Reconciliation Forms and URGENT INFORMATION RECALL NOTICE FOR INSTRUMENTS MODIFIED BY U.S. DISTRIBUTORS recall notifications were then delivered by DePuy Synthes Sales Consultants to the affected medical facilities. Instructions in the URGENT INFORMATION RECALL NOTICE FOR INSTRUMENTS MODIFIED BY U.S. DISTRIBUTORS provided to the medical facilities included the following: Please take the following actions: " Please immediately cease using the modified instruments identified in the attached Reconciliation Form. Your U.S. DePuy Synthes Sales consultant will work with your facility to locate and replace any affected instruments. " If your facility is using an instrument that was created or modified by a DePuy Distributor at the request of a surgeon, and it is not listed on the attached Reconciliation Form, please contact your DePuy Synthes Sales Consultant for an evaluation to determine if the instrument should be returned and replaced. WI-9956 | Rev 5 | Attachment B2 " Return Affected Instruments: o Medical facilities are to determine if any of the recalled instruments are still on hand by working with your U.S. DePuy Synthes Sales Consultant, and return affected devices immediately to their U.S. DePuy Synthes Sales Consultant or return them to DePuy Orthopaedics, Inc. for credit following normal purchasing procedures. o Note: These instruments may be on consignment at your facility. " Reconciliation Form: Complete the Reconciliation Form and return to your U.S. DePuy Synthes Sales Consultant or
Initiated
2017-01-06
Root cause
Vendor change control
Terminated
2019-05-15
Distribution
Nationwide Distribution to: AZ CA IA IL IN LA MA MD ME MI MN PA VA
Recall number
Z-2317-2017
Recall status
Terminated
CDRH record ID
151963
FDA record created
2017-06-08
Recall event number
76023

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗